According to the hospital: during an intervention for a cardiac arrest at 06.30 pm, the ecmo is primed and put in place without problem.The patient was connected.But during use, the medical staff observed a coherent rpm but with a flow at 0 l/min.The medical staff performed an external cardiac massage but the patient died.(b)(4).
|
(b)(4).A follow-up medwatch will be submitted when additional information becomes available.(b)(4).A service technician visited the hospital on 2019-01-09 and checked according to service (b)(4) the device.The technician could not find any failure at the device.The hospital informed us, that the patient was in a bad condition and they used the cardiohelp as a last solution.The event they complaint about, no flow, was caused according to the log-file analysis by a safety measure, the back flow prevention.As there was no malfunction no further action is needed.
|