BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. BD CONNECTA¿ STOPCOCK Back to Search Results
Catalog Number 394945
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/10/2018
Event Type  malfunction  
Manufacturer Narrative
Multiple lot numbers: there were multiple lot numbers reported to be involved.The information for each lot number is as follows: medical device lot #: 8226712, medical device expiration date: 2021-07-31, device manufacture date: 2018-09-24.Medical device lot #: 8180507, medical device expiration date: 2021-05-31, device manufacture date: 2018-08-30.(b)(6).A device evaluation and/or device history review is anticipated, but is not complete.Upon completion, a supplemental report will be filed.
 
Event Description
It was reported that bd connecta¿ stopcock injection valve leaks.No serious injury or medical intervention was reported.
 
Manufacturer Narrative
Investigation summary: a device history review was conducted for lot numbers 8226712 and 8180507.Our records show that this is the only instance of this issue occurring in either connect a batch.According to the sampling plan applied for product performance, each lot was accepted and released without defects being noted during the final assembly or visual inspections.Additionally, our engineers were able to duplicate this event through the leakage testing of the submitted device.A subsequent review of our manufacturing line determined that the most likely root cause for this event is an abnormality in the equipment responsible for tubing assembly.To prevent a reoccurrence of this event we have retrained our personnel and optimized our manufacturing process to monitor this issue more thoroughly.Based on investigation results to date, for leakage issue (in injection valve) root cause was associated to a bad tubing assembly by station 5 of equipment vh59, however, nogales site opened capa to perform an investigation.
 
Event Description
It was reported that bd connecta¿ stopcock injection valve leaks.No serious injury or medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD CONNECTA¿ STOPCOCK
Type of Device
STOPCOCK
Manufacturer (Section D)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MDR Report Key8196756
MDR Text Key131590031
Report Number9610847-2018-00460
Device Sequence Number1
Product Code FMG
Combination Product (y/n)N
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Type of Report Initial,Followup
Report Date 02/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/26/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number394945
Device Lot NumberSEE H.10
Date Manufacturer Received12/10/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-