SMITH & NEPHEW, INC. LGN AP FEM CUTTING BLK SZ 4; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Catalog Number 71434409 |
Device Problem
Break (1069)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/05/2018 |
Event Type
malfunction
|
Event Description
|
It was reported that during surgery the ball bearings that hold the valves collett in place came loose.Device broke in patient, all pieces accounted for.No delay or injury reported.No back up required, device use had been completed.
|
|
Manufacturer Narrative
|
The associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.Both ball plungers disassembled from the device and only one was returned with the device.The device was manufactured in 2017 and exhibits signs of moderate wear/usage.A dimensional inspection was performed and an out of tolerance feature was found that could have contributed to the failure.Corrective action is being addressed by the internal quality hold process.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.
|
|
Search Alerts/Recalls
|
|
|