BOSTON SCIENTIFIC CORPORATION UPHOLD LITE; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR PELVIC ORGAN PROLAPSE, TRANSVAGIN
|
Back to Search Results |
|
Model Number M0068318170 |
Device Problems
Break (1069); Use of Device Problem (1670)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 08/24/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).A visual examination of the returned capio slim device revealed that the cage was lifted up out of the capio slim head on one side.There was a hard, solid piece of debris located where the cage lifted out of the head which most likely caused damage to the cage.The uphold lite mesh assembly was not returned.The investigation concluded that the most probable cause for this complaint is adverse event related to procedure which indicates that the adverse event occurred during the procedure and the device had no influence on event.
|
|
Event Description
|
It was reported to boston scientific corporation that an uphold (tm) lite w/ capio slim was used during an anterior prolapse repair with uphold lite procedure performed on (b)(6) 2018.According to the complainant, during the procedure, the capio cage was damaged and unable to catch the dart.The procedure was completed with another uphold (tm) lite w/ capio slim.There were no serious injury or no adverse patients effects resulted from the procedure.The patient's condition was reported to be stable after the procedure.This event has been deemed reportable based on the investigation results: the capio cage was lifted up out of the capio slim head on one side.
|
|