• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION UPHOLD LITE; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR PELVIC ORGAN PROLAPSE, TRANSVAGIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION UPHOLD LITE; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR PELVIC ORGAN PROLAPSE, TRANSVAGIN Back to Search Results
Model Number M0068318170
Device Problems Break (1069); Use of Device Problem (1670)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/24/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A visual examination of the returned capio slim device revealed that the cage was lifted up out of the capio slim head on one side.There was a hard, solid piece of debris located where the cage lifted out of the head which most likely caused damage to the cage.The uphold lite mesh assembly was not returned.The investigation concluded that the most probable cause for this complaint is adverse event related to procedure which indicates that the adverse event occurred during the procedure and the device had no influence on event.
 
Event Description
It was reported to boston scientific corporation that an uphold (tm) lite w/ capio slim was used during an anterior prolapse repair with uphold lite procedure performed on (b)(6) 2018.According to the complainant, during the procedure, the capio cage was damaged and unable to catch the dart.The procedure was completed with another uphold (tm) lite w/ capio slim.There were no serious injury or no adverse patients effects resulted from the procedure.The patient's condition was reported to be stable after the procedure.This event has been deemed reportable based on the investigation results: the capio cage was lifted up out of the capio slim head on one side.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UPHOLD LITE
Type of Device
MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR PELVIC ORGAN PROLAPSE, TRANSVAGIN
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
FREUDENBERG MEDICAL
2301 centennial boulevard
jefferson IN 47130
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key8201273
MDR Text Key131587458
Report Number3005099803-2018-62230
Device Sequence Number1
Product Code OTP
UDI-Device Identifier08714729839200
UDI-Public08714729839200
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K122459
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 12/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/04/2020
Device Model NumberM0068318170
Device Catalogue Number72441
Device Lot Number0000061420
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/07/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/30/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/18/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age63 YR
-
-