SMITHS MEDICAL ASD, INC. CADD-PRIZM VIP SYSTEM; PUMP, INFUSION, PCA Back to Search Results
Catalog Number 21-8821-82
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
One cadd prizm vip pump was returned for analysis.Visual inspection performed and product found to be in good condition.Customer's complaint was not verified.The event log did not show the error.The motor will be replaced in service as a preventative measure.Use testing was performed.The problem source is unknown.
 
Event Description
Information was received that the cadd prizm vip pump had error code 10037.No reported adverse effects.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD-PRIZM VIP SYSTEM
Type of Device
PUMP, INFUSION, PCA
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
SMITHS MEDICAL ASD, INC.
1265 grey fox road
st. paul MN 55112
Manufacturer Contact
dave halverson
6000 nathan lane north
minneapolis, MN 55442
7633833310
MDR Report Key8205332
MDR Text Key131706475
Report Number3012307300-2018-08931
Device Sequence Number1
Product Code MEA
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 12/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number21-8821-82
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/04/2018
Date Manufacturer Received11/29/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-