• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. CADD-PRIZM VIP SYSTEM; PUMP, INFUSION, PCA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. CADD-PRIZM VIP SYSTEM; PUMP, INFUSION, PCA Back to Search Results
Model Number 6101
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Event Description
Information was received that the cadd prizm vip pump had error code 10037.No reported adverse effects.
 
Manufacturer Narrative
One cadd prizm pump was returned for analysis.Visual inspection performed and product found to be in good condition.Customer's complaint of error codes 10037 was verified.The event log did show the error code 10037.Service found the motor was intermittant.Service replaced the motor.Use testing was performed.The problem source is the faulty motor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD-PRIZM VIP SYSTEM
Type of Device
PUMP, INFUSION, PCA
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
MDR Report Key8205333
MDR Text Key131706483
Report Number3012307300-2018-08930
Device Sequence Number1
Product Code MEA
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Type of Report Initial,Followup
Report Date 02/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number6101
Device Catalogue Number21-8821-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2019
Date Manufacturer Received01/10/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-