• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD AIRVO2 HUMIDIFIER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD AIRVO2 HUMIDIFIER Back to Search Results
Model Number PT101
Device Problem No Audible Alarm (1019)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The complaint airvo is currently en route to fisher & paykel healthcare (b)(4) for evaluation.We will provide a follow up report upon completion of our investigation.
 
Event Description
A healthcare facility in (b)(6) reported that a speaker of pt101 airvo 2 humidifier was faulty.There was no patient involvement.
 
Manufacturer Narrative
Ps296056 method: the complaint airvo 2 humidifier was not returned to fisher & paykel healthcare (f&p) in new zealand.Our investigation is thus based on the information provided by the customer and our knowledge on the product.Results: the customer reported that the speaker of pt101 airvo 2 humidifier was not working properly.Previous investigation into this type of failure have identified that the problem is caused by a faulty speaker and electrical resistance testing has shown the speaker's resistance to be open circuit.Conclusion: as part of our ongoing product improvement initiatives, we have implemented a soak test for 100% testing of the speaker harness on the airvo production line, which identifies and discards any faulty speakers prior to assembly into the airvo.Additionally a new speaker unit has more recently been sourced from a different supplier.The subject airvo was manufactured prior to implementation of these measures.The airvo user manual states that the "airvo is for the treatment of spontaneously breathing patients who would benefit from receiving high-flow, warmed and humidified respiratory gases" and that "the unit is not intended for life support." the user manual warns the user: - prior to each patient use, ensure that the auditory alarm signal is audible by conducting the alarm system functionality check described in the alarms section the alarm system functionality check instructs the user on how to check the alarm function and states that "if either alarm signal is absent, do not use the unit.Contact your fisher & paykel healthcare representative.".
 
Event Description
A healthcare facility in australia reported that a speaker of pt101 airvo 2 humidifier was faulty.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIRVO2 HUMIDIFIER
Type of Device
AIRVO2 HUMIDIFIER
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
MDR Report Key8205415
MDR Text Key131940903
Report Number9611451-2018-01140
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
PMA/PMN Number
K131895
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPT101
Device Catalogue NumberPT101
Device Lot Number140714
Was Device Available for Evaluation? No
Date Manufacturer Received12/03/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-