• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEOMED, INC. NEOMED ENFIT SYRINGES; ENTERAL SYRINGES WITH ENTERNAL SPECIFIC CONNECTORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEOMED, INC. NEOMED ENFIT SYRINGES; ENTERAL SYRINGES WITH ENTERNAL SPECIFIC CONNECTORS Back to Search Results
Device Problems Incorrect Measurement (1383); Defective Device (2588)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  No Answer Provided  
Event Description
We recently went live with neomed enfit syringes, replacing all legacy oral syringes.We have had multiple adverse events using the 0.5ml syringe.Neomed¿s syringe cap snaps/clicks onto the syringe displacing the volume of air of the cap into in the syringe which creates variability in dose measurement.Despite doing advance testing, we did not catch this issue until live use.We also have had significant challenges with plunger movement with the cap on the syringe, in some cases up tp 0.1 ml of plunger movement.In other words, the technician fills the syringe and leaves for pharmacist checking.The pharmacist comes to check and the plunger will depress below the dose.More often 0.02 to 0.07ml of plunger movement after the dose is measured.This occurs more often with the oily medications like ferrous sulfate.These issues did not occur with our legacy syringes.After meeting with neomed¿s senior engineer, he showed us detailed cad drawings and a test model of a new syringe cap that will dispel air as you cap syringe.This was not revealed to us prior to go live, after further testing we noticed the controlled substances had less problems because the tamper evident cap screws on and therefore less air is forced into the syringe.We found no more than 0.02ml variation after the cap was placed with tamper evident caps.Neomed is giving us tamper evident caps for free ((b)(6) each) until the new cap is ready.All along neomed speaks of how many hospitals have launched with their product but not sharing the safety concerns in dose measurement.I understand they are making their way through a challenging design process but their transparency has been lacking.I watched a presentation by neomed yesterday intended for hospitals interested in learning about transition to enfit and they did not mention the cap issue.(b)(6), access number: (b)(4).Dosage form: injectable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEOMED ENFIT SYRINGES
Type of Device
ENTERAL SYRINGES WITH ENTERNAL SPECIFIC CONNECTORS
Manufacturer (Section D)
NEOMED, INC.
MDR Report Key8206288
MDR Text Key132492748
Report NumberMW5082682
Device Sequence Number1
Product Code PNR
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Type of Report Initial
Report Date 04/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/28/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator No Information
Was Device Evaluated by Manufacturer? No Information
Type of Device Usage N
Patient Sequence Number1
-
-