MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC BATTERY HANDPIECE/MODULAR FOR TRS; MOTOR, SURGICAL INSTRUMENT, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC BATTERY HANDPIECE/MODULAR FOR TRS; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 05.001.201
Device Problem Overheating of Device (1437)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/01/2018
Event Type  malfunction  
Manufacturer Narrative
As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported from (b)(6) that the battery handpiece/modular device became hot while being used together with the lid and power module device.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in (b)(6) of 2018.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Service review: a review of the service history record indicates that the device has been serviced within the last year for a service condition that is not relevant to the current reported condition.Device evaluation: this device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the reported condition was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation, it was determined that the etching was missing some parts, and the the yellow marking on the coupling nose was damaged.It was further determined that the device failed pretest for marking and labeling.It was determined that the issue was consistent with normal wear out from use.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY HANDPIECE/MODULAR FOR TRS
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key8206541
MDR Text Key131773742
Report Number8030965-2018-59328
Device Sequence Number1
Product Code GEY
UDI-Device Identifier7611819977815
UDI-Public(01)7611819977815(11)160927
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/31/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05.001.201
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/12/2018
Date Manufacturer Received02/11/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
LID DEVICE, POWER MODULE DEVICE
-
-