Catalog Number C-HSK-3043 |
Device Problem
Device Difficult to Setup or Prepare (1487)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/18/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).A lot history record review was completed for the reported product lot number.There were no ncmr¿s for the reported lot number.The device has not yet been returned to maquet cardiac surgery for evaluation.We are following up with the customer for the return of the device.A supplemental report will be submitted if the device is received.
|
|
Event Description
|
The hospital reported that during preparation for a coronary artery bypass procedure, hs iii proximal seal system 4.3mm seal did not load properly.A replacement device was used to complete the procedure.The hospital did not report any patient effects.
|
|
Event Description
|
The hospital reported that during preparation for a coronary artery bypass procedure, hs iii proximal seal system 4.3mm seal did not load properly.A replacement device was used to complete the procedure.The hospital did not report any patient effects.
|
|
Manufacturer Narrative
|
Internal complaint number: (b)(4).Autonumber: # (b)(4).The device was returned to the factory for evaluation.Signs of clinical use and no evidence of blood were observed.A visual inspection was conducted.The delivery device was returned inside the loading device.The seal could be seen from the loading device window.The delivery device was removed from inside the loading device.The seal and tension spring assembly remained inside the loading device.The slide lock was engaged.The plunger was not depressed on the delivery device.The seal was taken out from the loading device for inspection.No crack/delamination of seal was observed.The following measurements were taken; the inner delivery tube diameter was measured at.197 in.The outer diameter was measured at.221 in.The length of the delivery tube was measured at 2.49 in.The values recorded were within the tolerance specifications.Based upon the received condition of the device, the reported failure mode ¿fitting problem¿ was confirmed.
|
|
Search Alerts/Recalls
|