ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) CONFIRM; IMPLANTABLE CARDIAC MONITOR Back to Search Results
Model Number DM3500
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/18/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that device interrogation during procedure showed unexpected device condition.Device was not used and was replaced.Patient was stable.
 
Manufacturer Narrative
The reported field complaint of unexpected device condition was confirmed and was due to device exposure to low temperature.After stabilizing the device temperature, the device was interrogated several times without any anomaly.The device was tested on the bench and no anomalies were found.
 
Event Description
New information received notes that the unexpected device condition occurred due to exposure to cold condition.No patient was involved in the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONFIRM
Type of Device
IMPLANTABLE CARDIAC MONITOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key8207412
MDR Text Key131765687
Report Number2938836-2019-00008
Device Sequence Number1
Product Code MXC
UDI-Device Identifier05415067027320
UDI-Public05415067027320
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K163407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 04/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/01/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/14/2019
Device Model NumberDM3500
Device Catalogue NumberDM3500
Device Lot NumberS000061040
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/06/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/22/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/23/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age46 YR
-
-