Model Number 595000-001 |
Device Problem
Difficult to Insert (1316)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 12/20/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
This alleged failure mode poses a low risk to a patient because the issue was observed when the freedom driver was not supporting a patient.The freedom driver has been returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
The freedom driver was not supporting a patient.The supplier, a syncardia authorized warehouse, reported needing more force than usual to insert onboard battery into the right freedom driver battery well.
|
|
Manufacturer Narrative
|
The customer reported difficulty inserting the onboard battery into the right battery well of the freedom driver was confirmed during investigation testing of the driver, where the battery release button in the right battery well stayed partially depressed thereby making it difficult to insert an onboard battery into the well.Despite the difficulty in battery insertion, the driver passed all functional testing requirements.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
|
|