Follow-up #1 and final report submitted to update sections d.3, g.1, g.4, g.7, h.2, h.3, h.6, h.10 and h.11 based on the results of investigation.Reported event: we opened the power to set up the room and the first power (42031116) would not accept the reamer adaptor.We swapped it out with a second power that failed during the surgery.The next day i tested both powers to see if they would accept the reamer adaptor and the one that failed would, but the other would not.So i have one that has no power and another that does not work.=15 minutes surgical delay.Product evaluation and results: the product was not evaluated as the product was unavailable for inspection.Capa 2127499 product history review: device history records indicate 25 devices were manufactured under lot k09gu and 11 devices were inspected and accepted into final stock on 5/31/17.No non-conformances were identified during inspection.14 other devices were inspected and 13 devices were accepted into final stock on 05/10/2017.A review of qt-17-05-0043 revealed that the non-conformance is not related to the failure alleged in this compliant.Complaint history review: a review of complaints in catsweb and trackwise related to p/n 209063, prodex lot k09gu shows no additional complaints related to the failure in this investigation.Conclusions: the alleged failure mode could not be confirmed as the product was not available for inspection.No additional investigation or specific actions are required.Corrective action/preventive action: a review of stryker¿s nc/capa database indicated there have been nc 1414517 and capa 1450904 associated with the product and failure mode reported in this event.Product was not available for evaluation.
|