Catalog Number 8065990794 |
Device Problem
Device Sensing Problem (2917)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The device history record (dhr) for the device has been reviewed.The associated device was released based on company's acceptance criteria.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A customer reported that centration was different in the microscope versus with the eye tracker.Patients outcome after procedure was good.Additional information has been requested.
|
|
Manufacturer Narrative
|
During technical onsite visit, the field service engineer (fse) completed system verification.The system was operating within specifications.No problem detected.No technical root cause was identified as the product was found to be within specifications during onsite visit.The root cause could not be determined conclusively.The manufacturer internal reference number is: (b)(4).
|
|
Search Alerts/Recalls
|