DEPUY SYNTHES PRODUCTS LLC ELECTRIC PEN DRIVE 60,000 RPM; MOTOR, SURGICAL INSTRUMENT, AC-POWERED
|
Back to Search Results |
|
Catalog Number 05.001.010 |
Device Problem
Overheating of Device (1437)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 10/16/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
The manufacturing location is currently not available.Service review: a review of the service history record indicates that the device has been serviced within the last year for a service condition that is not relevant to the current reported condition.Device evaluation: the actual device was returned for evaluation.The device was evaluated and the reported condition was confirmed.A visual and functional assessment was performed on the device which found that the device started to run and then it stopped when it was getting warm.It was determined that the device failed for check function and direction of rotation, check function with test hand switch and check function with foot pedal.During service/repair, it was observed that the internal electronic components were damaged.The assignable root cause was determined to be due to normal wear out from use.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Event Description
|
It was reported that during pre-surgery or set up, it was observed that the electric pen drive device had no power and was not working.During in-house engineering evaluation, it was observed that the device started to run and then it stopped when it was getting warm.The event was not related to surgery.There was no patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Search Alerts/Recalls
|
|
|