Catalog Number 00882100100
Device Problem Electrical Shorting (2926)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/02/2019
Event Type  malfunction  
Manufacturer Narrative
This event is recorded with zimmer biomet under (b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that the device power cord was cut.The event occurred during reprocessing there was no harm and no delay reported.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.The following sections were updated/corrected: b4, d4, d10, g4, g7, h2, h3, h4, h6, h10 the device history record and previous repair record for zimmer electric dermatome serial number (b)(4) was reviewed and noted no related non-conformances, requests for deviation (rfd), change notices (cn), or any other issues with manufacturing.The record reviews found no issues with the device and all verifications, inspections, and tests were successfully completed.The reported event was confirmed by the service technician who performed the evaluation and repair.On 4 january 2019, it was reported from trimedx that a dermatome had its power cord cut.The customer returned a zimmer electric dermatome, serial number (b)(4), for evaluation.Evaluation of the device on 14 january 2019 noted that the power cord was damaged, but the device was still within side to side calibration and motor speed specifications.The control bar was also found to be in the correct position.Repair of the device occurred the same day and involved replacing the power cord, motor, power switch, and multiple bearings.The technician then tested and verified that the dermatome was functioning as intended and the device was returned to the customer without further incident.The device was tested, inspected, and repaired.Reference number (b)(4) on 4 january 2019.While the service technician confirmed that the power cord had been damaged, it cannot be determined from the information provided as to what transpired such that the power cord became damaged.Therefore, a specific root cause of the cut power cord cannot be determined.The investigation was based on the information that was provided initially and any information that was obtained throughout the follow-up process.
 
Event Description
No additional event information was received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDPIECE, ELECTRIC DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
MDR Report Key8223843
MDR Text Key132359666
Report Number0001526350-2019-00008
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 03/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/07/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number00882100100
Device Lot Number63138521
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/08/2019
Was the Report Sent to FDA? No
Date Manufacturer Received03/19/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-