• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PS X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PS X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5532-G-619
Device Problems Material Deformation (2976); Naturally Worn (2988)
Patient Problems Cyst(s) (1800); Injury (2348); Inadequate Osseointegration (2646); Fibrosis (3167)
Event Date 05/14/2012
Event Type  Injury  
Manufacturer Narrative
Review of the device history records indicate that devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Device not available.
 
Event Description
Letter received from patient's lawyers advising the stryker ts triathlon knee replacement is "defective and materially contributed to on going left lower limb problems including the development of a pseudo-tumour".Patient is reports left knee problems following surgery on the (b)(6) 2012.Please see communication log for details regarding additional information.As per op report: comments triathlon ts cemented revision knee from an (b)(6) year old male revised after 4 years due to osteolysis and alval (pseudotumour) is presented for analysis.High activity level and medial bone loss around the tibial component are considered contributory factor to failure of the device.Femoral component: a combination of bone and fibrous tissue onto cement mantle indicating a well fixed components - numerous longitudinal scratches on bearing surface.Threaded boss covered with organic material and showing wear on the surface in contact with stem extender (fig.1).Femoral stem extender: corrosion and wear in the base of threaded trunnion from uneven engagement with femoral component boss covered with organic material (fig.2).Femoral stem: wear in the base of threaded trunnion from uneven engagement with femoral.Extender tibia baseplate: polished cement surface, indicative of micromotion, postero-medially.Bone tissue still attached to the cement mantle, anteriorly.Severe fretting corrosion of the surface of the boss as well as corrosion deposits in the threaded boss in contact with stem trunnion.Tibial stem: component was loose and removed from the baseplate manually.Gross fretting and corrosion in the trunnion: stem surface (inset, fig.4), and in the trunnion threads (fig.5) polyethylene (pe) insert: slight wear on the bearing surface, mostly scratching and burnishing.Burnishing around the ps post, posteriorly and media-lateral regions from contact with the femoral component.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON PS X3 TIBIAL INSERT
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
juana hiciano
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key8229280
MDR Text Key132482674
Report Number0002249697-2019-00117
Device Sequence Number1
Product Code MBH
UDI-Device Identifier07613327046380
UDI-Public07613327046380
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
K141056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Reporter Occupation Other
Type of Report Initial
Report Date 01/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Device Operator No Information
Device Expiration Date10/31/2013
Device Catalogue Number5532-G-619
Device Lot NumberXE7MNE
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/11/2018
Initial Date FDA Received01/08/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/28/2008
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age58 YR
-
-