MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910 PFC*CALIBRATED PAT CUT GDE; KNEE INSTRUMENT : PATELLA HANDLE/CLAMPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910 PFC*CALIBRATED PAT CUT GDE; KNEE INSTRUMENT : PATELLA HANDLE/CLAMPS Back to Search Results
Catalog Number 865034
Device Problems Incorrect Measurement (1383); Defective Device (2588)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/11/2018
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
This is for a j and j specialist 2 patella instrument clamp.The surgeon measured to leave 14mm of residual bone and instead it left around 10-11mm of patella after the resection.The clamp was reviewed at the end of the case and indeed it was set for 14mm.The surgeon ended up sawing a small wafer of bone from the resected patella with ethibond suture onto the patella before implanting.The surgeon wanted the clamp looked into and following the case the clamp was taken out of the tray and put a repair tag on it to be washed up and sent back to johnson and johnson.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint (b)(4).Investigation summary :examination of the returned instrument confirmed the reported event.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
KNEE INSTRUMENT : PATELLA HANDLE/CLAMPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8235204
MDR Text Key132785097
Report Number1818910-2019-80408
Device Sequence Number1
Product Code HWT
UDI-Device Identifier10603295216599
UDI-Public10603295216599
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 12/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/10/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number865034
Device Lot NumberJO604
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2019
Date Manufacturer Received02/06/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age73 YR
-
-