PENUMBRA, INC. INDIGO SYSTEM ASPIRATION CATHETER 8; DXE Back to Search Results
Catalog Number CAT8XTORQ115
Device Problems Device Damaged Prior to Use (2284); Deformation Due to Compressive Stress (2889)
Patient Problem No Patient Involvement (2645)
Event Date 12/13/2018
Event Type  malfunction  
Manufacturer Narrative
The device has been returned and the investigation results are pending.A follow up mdr will be submitted upon completion of the device investigation.
 
Event Description
During preparation for a medical procedure, the hospital staff found that the indigo system aspiration catheter 8 (cat8) had been crushed at the middle of the catheter, before removal from the packaging.The damage to the cat8 was found prior to use and therefore it was not used in the procedure.The procedure was completed using another cat8.
 
Manufacturer Narrative
Results: the cat8 was ovalized from approximately 48.0 ¿ 49.0 cm from the hub.Conclusions: evaluation of the returned cat8 confirmed an ovalized device.The reported complaint mentioned that the damage occurred before removal of the device from its packaging.This damage was likely a result of improper handling during transit or during storage of the device.The root cause of the reported complaint could not be determined.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INDIGO SYSTEM ASPIRATION CATHETER 8
Type of Device
DXE
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key8241909
MDR Text Key132916614
Report Number3005168196-2019-00076
Device Sequence Number1
Product Code DXE
UDI-Device Identifier00814548016290
UDI-Public00814548016290
Combination Product (y/n)Y
PMA/PMN Number
K160533
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 01/01/2005,12/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/11/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/12/2021
Device Catalogue NumberCAT8XTORQ115
Device Lot NumberF85199
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2019
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Date Manufacturer Received02/26/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-