Catalog Number 010000661
Device Problem Delivered as Unsterile Product (1421)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/16/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Complaint sample was evaluated and the reported event was confirmed.Visual inspection of the returned product and review with the manufacturing site identified that the vacuum seal of the shell and screw are inadequate per specifications.Device history record was reviewed and no discrepancies were found.The likely condition of the product when it left zimmer biomet was non-conforming.The reported event occurred during the manufacturing process.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that the product was found to be improperly sealed.As the sterile barrier was damaged, a new product was used to complete the procedure without delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 PPS LTD ACET SHELL 48C
Type of Device
HIP, PROSTHESIS
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key8244893
MDR Text Key132964441
Report Number0001825034-2019-00155
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/14/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number010000661
Device Lot Number6062455
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/22/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/18/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/31/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-