• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK ENDOSCOPY CAPTURA PRO¿ BIOPSY FORCEPS WITH SPIKE; PTS ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED, EXEMPT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK ENDOSCOPY CAPTURA PRO¿ BIOPSY FORCEPS WITH SPIKE; PTS ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED, EXEMPT Back to Search Results
Catalog Number BF-2.2S-230SP-20
Device Problem Dull, Blunt (2407)
Patient Problems Tissue Damage (2104); No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.A review of the device history record could not be conducted because the lot number was not provided.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The size of tissue sample or biopsy to be excised can be controlled by how the user handles the forceps.If excessive pressure was applied during advancement into the tissue or during handle manipulation, this could have contributed to the reported observation.The instructions for use direct the user to advance the forceps into the tissue at the desired biopsy site.Then the user is instructed to close the forceps around the tissue while using slight pressure on the handle.The user is instructed to maintain gentle handle pressure to keep the cups closed and gently withdraw the forceps from the site.The instructions for use (ifu) warns: "these single-use forceps should only be used to biopsy tissue where possible bleeding or hemorrhage will not present a danger for patients.Adequate plans for management of potential bleeding or hemorrhage and appropriate airway management should be in place." the ifu lists potential complications as: "those associated with gastrointestinal endoscopy include but are not limited to: perforation, bleeding or hemorrhage, aspiration, fever, infection, allergic reaction to medication, hypotension, respiratory depression or arrest, cardiac arrhythmia or arrest." prior to distribution, all captura pro disposable biopsy forceps are subjected to a visual inspection to ensure device integrity.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event Description
During multiple gastrointestinal procedures, the physician used multiple captura pro biopsy forceps with spike devices.When taking a biopsy, the tissue tears.This occurs in lower and upper gastrointestinal procedures.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTURA PRO¿ BIOPSY FORCEPS WITH SPIKE
Type of Device
PTS ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED, EXEMPT
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
scottie fariole
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key8250692
MDR Text Key133284230
Report Number1037905-2019-00020
Device Sequence Number1
Product Code PTS
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 12/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/15/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberBF-2.2S-230SP-20
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/21/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-