• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT POINT OF CARE I-STAT CG4+ CARTRIDGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT POINT OF CARE I-STAT CG4+ CARTRIDGE Back to Search Results
Catalog Number 03P85-50
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problems No Information (3190); No Code Available (3191)
Event Date 12/26/2018
Event Type  malfunction  
Manufacturer Narrative
Apoc incident # (b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
 
Event Description
On (b)(6) 2018, abbott point of care was contacted by a customer regarding i-stat cg4+ cartridges (lot# unk) that yielded suspected discrepant pco2 results during a study of hypothermic babies.There was no patient information available at the time of this report.The customer stated that they are not live for testing at this time and that the results were not used to treat the patient.(b)(6).Test and collection times not available.There are no injuries associated with this event.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on the limited information available that suggests the product was not performing within the variability of the assay.The investigation is underway.
 
Manufacturer Narrative
Apoc incident #: (b)(4).The investigation was completed on 03/26/2019.A review of the device history record (dhr) confirmed the cartridge lot met finished goods (fg) release criteria.Retained testing met the acceptance criteria outlined in appendix 1 of q04.01.003 rev.Ac (product complaint level 2 and level 3 investigation procedure).No deficiency has been determined for cg4+ lot d18246a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT CG4+ CARTRIDGE
Type of Device
CG4+ CARTRIDGE
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
MDR Report Key8253294
MDR Text Key134579781
Report Number2245578-2019-00011
Device Sequence Number1
Product Code KHP
UDI-Device Identifier01005474902266
UDI-Public1005474902266
Combination Product (y/n)N
PMA/PMN Number
K982071
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 04/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/16/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03P85-50
Was Device Available for Evaluation? No
Date Manufacturer Received03/26/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-