• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PLEXUS MANUFACTURING SDN. BHD; PULSE-GENERATOR, PACEMAKER, EXTERNAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PLEXUS MANUFACTURING SDN. BHD; PULSE-GENERATOR, PACEMAKER, EXTERNAL Back to Search Results
Model Number 5392
Device Problem Material Integrity Problem (2978)
Patient Problem No Patient Involvement (2645)
Event Date 01/09/2019
Event Type  malfunction  
Manufacturer Narrative
This device was reported as included in the field action noted and returned product investigation found the device performed as described in the field action.Product event summary: manufacturer¿s analysis found that the external pulse generator (epg) had a broken and dented upper case, broken lower case, pinched output connector wire, pinched display wires with exposed insulation, scratched keypad, contaminated encoder assembly and knob, pinched main seal, and two contaminated output connector nuts.All found defective parts were replaced and all other identified issues were resolved.The device passed all final functional tests.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the external pulse generator (epg) originally returned due to a field corrective action (fca) subsequently tested out of specification during manufacturer¿s analysis.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY   11900
Manufacturer Contact
lisa robertson
8200 coral sea st ne
mounds view, MN 55112
7635262723
MDR Report Key8255653
MDR Text Key133327237
Report Number3004593495-2019-00055
Device Sequence Number1
Product Code DTE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial
Report Date 01/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/16/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/26/2018
Date Manufacturer Received01/09/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/14/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-1356-2015
Patient Sequence Number1
-
-