MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR FEM COMP #5 R-CEM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR FEM COMP #5 R-CEM; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5510F502
Device Problem Packaging Problem (3007)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/20/2018
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of the investigation.
 
Event Description
Customer reported that when the opened a femoral triathlon implant, they found the inner packaging was damaged (split).Another implant was used to complete the case, and no surgical delays.
 
Manufacturer Narrative
An event regarding pack damage involving triathlon femoral component was reported.The event was confirmed.Method & results: device evaluation and results: visual inspection: visual inspection of the returned device indicate packaging damage on the outer blister.Inner blister appears to be intact.Minimal damage can be seen on the outer box.Medical records received and evaluation: no medical records were received for review with a clinical consultant device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot.Conclusion: the investigation concluded that the packaging damage was most likely caused by incorrect handling prior to the opening packaging whereby the carton may have been compressed and/or dropped from a height causing the damage to the carton.No further investigation for this event is required at this time.
 
Event Description
Customer reported that when the opened a femoral triathlon implant, they found the inner packaging was damaged (split).Another implant was used to complete the case, and no surgical delays.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON CR FEM COMP #5 R-CEM
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8257872
MDR Text Key133537551
Report Number0002249697-2019-00219
Device Sequence Number1
Product Code MBH
UDI-Device Identifier07613327040302
UDI-Public07613327040302
Combination Product (y/n)N
PMA/PMN Number
K141056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Type of Report Initial,Followup
Report Date 04/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/17/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2023
Device Catalogue Number5510F502
Device Lot NumberD4L6A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/16/2019
Date Manufacturer Received03/08/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-