MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION TRAPEZOID RX; LITHOTRIPTOR, BILIARY MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION TRAPEZOID RX; LITHOTRIPTOR, BILIARY MECHANICAL Back to Search Results
Model Number M00510890
Device Problems Deformation Due to Compressive Stress (2889); Difficult to Open or Close (2921); Material Deformation (2976); Material Twisted/Bent (2981)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/26/2018
Event Type  malfunction  
Manufacturer Narrative
Patient's exact age is unknown; however it was reported that the patient was over the age of 18.(b)(4).The device is not expected to be returned for analysis.If any further relevant information is received, a supplemental mdr will be filed.
 
Event Description
It was reported to boston scientific corporation that a trapezoid rx lithotripter basket was used in the common bile duct (cbd) during a biliary stone retrieval procedure performed on (b)(6) 2018.According to the complainant, after being inserted into the cbd during the procedure, an attempt to grasp the stone for extraction was performed; however, when trying to open and close the basket two times, the sheath detached from the black heat shrink.The basket was removed and the procedure was completed with a different device.Additionally, a photograph included from the customer shows that the sidecar rx tunnel was pushed back and the coil was bent.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAPEZOID RX
Type of Device
LITHOTRIPTOR, BILIARY MECHANICAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
,
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
,
marlborough, MA 01752
5086834015
MDR Report Key8268839
MDR Text Key133846823
Report Number3005099803-2019-00065
Device Sequence Number1
Product Code LQC
UDI-Device Identifier08714729296409
UDI-Public08714729296409
Combination Product (y/n)N
Reporter Country CodeIR
PMA/PMN Number
K040447
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 01/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/22/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/03/2019
Device Model NumberM00510890
Device Catalogue Number1089
Device Lot Number0022072960
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/27/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/03/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial