It was reported to boston scientific corporation that a trapezoid rx lithotripter basket was used in the common bile duct (cbd) during a biliary stone retrieval procedure performed on (b)(6) 2018.According to the complainant, after being inserted into the cbd during the procedure, an attempt to grasp the stone for extraction was performed; however, when trying to open and close the basket two times, the sheath detached from the black heat shrink.The basket was removed and the procedure was completed with a different device.Additionally, a photograph included from the customer shows that the sidecar rx tunnel was pushed back and the coil was bent.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
|