It was reported that this right ventricular (rv) lead exhibited low amplitude, noise, oversensing, pacing inhibition, and loss of capture.The patient experienced chest and pocket stimulation.It was noted the impedances had decreased from 700 ohms at implant to 601 ohms.Fluoroscopy was performed and the lead was still in the rv apex.Subsequently, a revision procedure was performed and the rv lead was explanted and replaced.There were two cuts noted in the lead body.No additional adverse patient effects were reported.
|
Upon receipt at our post market quality assurance laboratory, visual inspection revealed several cuts in the insulation through to all lumens.This type of damage is consistent with damage induced during the explant procedure.Testing was completed to assess lead electrical performance and inner insulation integrity.Measurements throughout these tests were within normal limits.Laboratory analysis did not identify any lead characteristics that would have caused or contributed to the reported low amplitude, loss of capture, low impedances, noise, oversensing, and pacing inhibition.
|