Model Number MC1VR01US |
Device Problem
Battery Problem (2885)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/03/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the leadless implantable pulse generator (ipg) experienced premature battery depletion, and the physician was concerned that the battery could fail without warning.The device was inactivated, and a new leadless ipg was implanted.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Product event summary: the device was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory showed the battery is before/approaching the indicator signifying that it the time for device replacement.If information is provided in the future, a supplemental report will be issued.
|
|