(b)(4).Device evaluation by manufacturer: a field service engineer (fse) visited the customer's site to address the event.Fse verified the issue on the error log and was also able to reproduce the error.Fse continued evaluation and found tube from line filter occluded.Fse replaced the occluded tubing.Customer replaced the column and fse successfully completed calibration of instrument.Fse also completed quality control and calibration runs; all results were in acceptable range.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(4) through aware date.There were no similar complaints identified during the searched period.The g8 variant analysis mode operator's manual under chapter 6, troubleshooting, states the following: 6.3 error messages: when consulting with technical support about a problem, please note the error message and error number.In addition, if you follow the suggested solutions in this section and are still unable to resolve the error, or if you encounter an error message that is not noted, contact technical support.General error messages: with these errors, the assay stops and the analyzer immediately enters stand-by state.100 pressure high is generated when the pump pressure exceeded the upper limit (15 mpa) set in the pres high parameters.Solution: when the filter or column replacement period has been exceeded, first replace the filter or column.If the pressure is still high, remove the inlet and outlet flow line around the column and filter, and determine which part is the cause of the high pressure.Then, contact a technical support representative.If the pressure displayed on the screen is: greater than the pressure on the column inspection report + 4 mpa, then replace the filter.Less than the pressure on the column inspection report, then proceed with priming the column.The most probable cause of the reported event was due to occluded tubing from line filter to preheater.
|
A customer reported that they were getting error 100 pressure high after installing a new column lot df1027n on the g8 instrument.The customer stated that the pressure reading was 13.1mpa and the upper limit of pressure for this column is 12.0mpa.Flow factor is 1.14 and sa1c retention time is 0.59.Technical support specialist (tss) ordered new column.The customer received the new column and installed on the instrument, but pressure is still high at 15.5mpa.Tss instructed the customer to remove the tubing from the prefilter and turn the pump on, the pressure reading decreased to 0.0mpa.The customer reconnected the tubing to the prefilter and pressure reading kept fluctuating to 15.0, 8.7, 5.0, 12.0mpa.The customer checked the filter and confirmed it was installed properly.A field service engineer (fse) was dispatched to address the reported event, which resulted in delay in reporting of patient results for hemoglobin a1c (hba1c).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|