ST. JUDE MEDICAL ADVISOR¿ HD GRID MAPPING CATHETER, SENSOR ENABLED¿; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
|
Back to Search Results |
|
Model Number D-AVHD-DF16 |
Device Problems
Difficult to Remove (1528); Material Separation (1562)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/23/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
One image was submitted for evaluation.No devices were returned.Based solely on the image submitted, paddle spline shafts appeared to be fractured and detached from a catheter shaft.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information provided, the cause of the reported damaged occurred during removal.
|
|
Event Description
|
During a procedure, resistance was felt when withdrawing the advisor hd grid catheter.Fluoroscopy was used and found that the catheter had become entangled with the other diagnostic catheters in the low groin area.The hd grid catheter was removed with force, which removed the outer covering of the catheter.All catheters were able to be removed with no adverse consequences to the patient.The procedure was then continued as intended.
|
|