Model Number D-AVHD-DF16
Device Problems Difficult to Remove (1528); Material Separation (1562)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/23/2019
Event Type  malfunction  
Manufacturer Narrative
One image was submitted for evaluation.No devices were returned.Based solely on the image submitted, paddle spline shafts appeared to be fractured and detached from a catheter shaft.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information provided, the cause of the reported damaged occurred during removal.
 
Event Description
During a procedure, resistance was felt when withdrawing the advisor hd grid catheter.Fluoroscopy was used and found that the catheter had become entangled with the other diagnostic catheters in the low groin area.The hd grid catheter was removed with force, which removed the outer covering of the catheter.All catheters were able to be removed with no adverse consequences to the patient.The procedure was then continued as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVISOR¿ HD GRID MAPPING CATHETER, SENSOR ENABLED¿
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
stephanie o' sullivan
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key8288320
MDR Text Key134446979
Report Number3005334138-2019-00040
Device Sequence Number1
Product Code DRF
UDI-Device Identifier05415067028198
UDI-Public05415067028198
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/30/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2020
Device Model NumberD-AVHD-DF16
Device Lot Number6750817
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/23/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/26/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age22 YR
Patient Weight69
-
-