MEDTRONIC SOFAMOR DANEK USA, INC X-STOP PEEK INTERSPINOUS SPACER; PROSTHESIS, SPINOUS PROCESS SPACER/PLATE
|
Back to Search Results |
|
Catalog Number 1-3014 |
Device Problem
Break (1069)
|
Patient Problem
Disability (2371)
|
Event Type
Injury
|
Manufacturer Narrative
|
This part is not approved for use in the united states; however a like device catalog # 1-3212, pma #p040001, udi # (b)(4) was cleared in the united states.Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported via medwatch that on an unknown date, post-op, patient had anterior slip due to cage breakage.Patient underwent fusion at l5/l4/l3 in which cage was fitted.After 4 months, the cage broke apart which had made patient bedridden and the patient had to take several painkilling injections.As an emergency, the implanted device was taken out.Patient was put on morphine till now.
|
|
Search Alerts/Recalls
|
|
|