Ous mdr - after implantation, it was reported that, due to an exit block in the rv, no stimulation threshold and no impedance in the ventricle could be determined during follow-up.
|
After the device was received, the pacemaker status was interrogated and the memory was analyzed.The pacemaker was visually inspected and the connection system was examined.The screw and the spring element of the lead connections were in working order.Leads inserted for test purposes were easily accessed, low ohm, and reliably contacted.The bore dimension was also within the dimensions set forth in the standard.The therapy capabilities of the pacemaker were checked.The anti-bradycardia output signal reflected the values programmed and the device signal sensing was in working order.With respect to device function, the pacemaker was within specifications.In addition, a long-term pacing test was performed.No problems were found with the pacing function.The device was not found to be compromised in its ability to deliver therapy.In summary, the device was fully functional during analysis, as was within specifications with respect to the connection system and electronics.There was no material or manufacturing defect present.
|