• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MASIMO - 40 PARKER RAD-8, HORIZONTAL; OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MASIMO - 40 PARKER RAD-8, HORIZONTAL; OXIMETER Back to Search Results
Model Number 22042
Device Problem No Audible Alarm (1019)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/26/2018
Event Type  malfunction  
Manufacturer Narrative
The returned device was evaluated.Visual inspection upon receiving revealed no external damage.The device was able to power on via battery and ac power and was able to obtain spo2 and pulse rate readings.The unit was found to visually alarm during alert conditions.Audio was silent; no sound was emitted from front or bottom speaker, due to failure of r62 component on the system board which was caused by an open circuit.A service history record review reveals that this unit was in the field for over two (2) years with no previous reported issues prior to this reported event.
 
Event Description
The customer reported the rad-8 speaker is not working.No patient impact or consequences were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAD-8, HORIZONTAL
Type of Device
OXIMETER
Manufacturer (Section D)
MASIMO - 40 PARKER
40 parker
irvine CA 92618 1604
Manufacturer (Section G)
MASIMO - MEXICALI
industrial vallera de mexicali calzada del oro, no.2001
mexicali, baja california 21600
MX   21600
Manufacturer Contact
erica kline
52 discovery
irvine, CA 92618-1604
9492977863
MDR Report Key8299536
MDR Text Key134875800
Report Number2031172-2019-00057
Device Sequence Number1
Product Code DQA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K092838
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Biomedical Engineer
Remedial Action Other
Type of Report Initial
Report Date 12/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/01/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number22042
Device Catalogue Number9190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/10/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/22/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-