NUVASIVE, INC. MAXCESS; STERILIZATION WRAP CONTAINERS, TRAYS, CASSETTES & OTHER ACCESSORIES Back to Search Results
Device Problems Thermal Decomposition of Device (1071); Smoking (1585)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/07/2019
Event Type  malfunction  
Event Description
The light source used (from the nuvasive maxcess surgical access kit) was plugged into the light box by the rn and then the staff noticed smoking and a burning smell coming from the light box at the point where the light source was plugged in.At this point, i am back from break and the nuvasive representative was bringing in another kit to use the light source out of it.The rn had already removed the light box from the room and brought in a different one.When she went to plug in the second light source the other representative said to hold off plugging it in because some light boxes have caused the metal end that gets inserted into the box to overheat "and there's an adaptor we are getting for you, so hold off a sec".They gave us the adaptor, we plugged it in and no further problems were noted.Both the light box and the cord were sent to biomed.Operating room leadership review: investigation summary: nurse in the room notified me that the light source was smoking from the port where the light cord was plugged in.Light cord was part of a kit supplied by nuvasive.Nuvasive rep told staff in the rooms that this occurs occasionally and that an adaptor was needed for our light source.Nurse was instructed to disconnect the light cord from the light source and take both items out of use immediately, then to save all packaging and the cord itself and put in a work order for the light source.Light source, cord, and packaging were taken to the biomed office for evaluation.Why did this happen? issue with light cord.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAXCESS
Type of Device
STERILIZATION WRAP CONTAINERS, TRAYS, CASSETTES & OTHER ACCESSORIES
Manufacturer (Section D)
NUVASIVE, INC.
101 enterprise ste 100
aliso viejo CA 92656
MDR Report Key8309976
MDR Text Key135158244
Report Number8309976
Device Sequence Number1
Product Code KCT
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/21/2019,01/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/06/2019
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/21/2019
Event Location Hospital
Date Report to Manufacturer02/06/2019
Date Manufacturer Received03/25/2019
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age14965 DA
-
-