• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INTELLAMAP ORION HIGH RESOLUTION MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INTELLAMAP ORION HIGH RESOLUTION MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number M004RC64S0
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/30/2018
Event Type  malfunction  
Event Description
Event reportable upon analysis completed on 09jan2019.It was reported that during a atrial flutter ablation procedure a magnetic sensor error was displayed when the intellamap orion catheter was connected.The procedure was completed using another catheter.No patient complications were reported.However analysis of the device revealed a separation of the sensor array.Visual inspection revealed a separation at the sensor array.The electrical test was performed and the device failed the test.X-ray shows that at the point of the misaligned layers of the sensor array the wires are open.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLAMAP ORION HIGH RESOLUTION MAPPING CATHETER
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
302 parkway
global park, la aurora,
heredia
CS  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330,
saint paul, MN 55112
6515826168
MDR Report Key8310930
MDR Text Key135189502
Report Number2134265-2019-00860
Device Sequence Number1
Product Code DRF
UDI-Device Identifier08714729841968
UDI-Public08714729841968
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K143481
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 02/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/06/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/08/2019
Device Model NumberM004RC64S0
Device Catalogue Number87035
Device Lot Number0022227302
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/20/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/09/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/08/2018
Is the Device Single Use? Yes
Type of Device Usage N
Patient Sequence Number1
-
-