Catalog Number CATD |
Device Problems
Device Damaged Prior to Use (2284); Material Deformation (2976)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 01/08/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device has been returned and the investigation results are pending.A follow up mdr will be submitted upon completion of the device investigation.Additional 510(k)#s that also apply to this complaint: k160533, k161523.
|
|
Event Description
|
During preparation for a medical procedure, the hospital staff was inspecting an indigo system catd aspiration catheter (catd) and found a kink near the distal end of the catd.The damage to the catd was found prior to use and therefore it was not used in the procedure.The procedure was completed using another catd.
|
|
Manufacturer Narrative
|
Results: the returned catd was kinked at approximately 41.0 cm from the hub.Conclusions: evaluation of the returned catd confirmed a kink on the distal shaft of the catheter.If the catheter is forcefully removed from the packaging tray at extreme angles or otherwise mishandled during preparation, damage such as a kink may occur.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
|
Search Alerts/Recalls
|