The patient's generator and lead were explanted and returned for product analysis.The patient's lead was replaced due to incompatibility with the desired replacement generator.The lead was returned in three pieces and the portion of the lead including the electrode array was not returned for analysis, and therefore commentary cannot be made on that portion of the lead.An abraded opening was identified in the outer and inner tubing, which was likely the cause of the dried remnants of what appear to have once been body fluid inside of the outer and inner silicone tubing.With exception of the noted abraded opening, the condition of the returned lead portions is consistent with the conditions typically seen after explant procedure with no other anomalies noted.Product analysis was completed on the explanted generator.The generator was monitored for 24 hours in a simulated body temperature environment and the generator provided the expected level of output current for the entirety of the monitoring period.The generator performed according to functional specifications with no anomalies identified.No further relevant information has been received to date.
|