MAUDE Adverse Event Report: SMITHS MEDICAL ASD; INC. PORTEX® SINGLE SHOT EPIDURAL TRAYS; ANESTHESIA CONDUCTION KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD; INC. PORTEX® SINGLE SHOT EPIDURAL TRAYS; ANESTHESIA CONDUCTION KIT Back to Search Results
Catalog Number NEPI-NLD-15597-20
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/15/2018
Event Type  Injury  
Event Description
Information was received indicating that the bupivicain to a smiths medical portex® single shot epidural tray failed to work and the patient was converted to general anesthesia.There were no reported adverse patient effects.
 
Manufacturer Narrative
One epidural tray was returned for evaluation.Visual inspection of the returned device has found it to be in good physical condition.The drug liquid was found to be clear (colorless) with no particulates visible inside the vials.Assay testing, performed by the supplier prior to release of bupivacaine lot # 87-159-ev was within specification.Review of the sterilization certificate for the tray lot number 3669767 (po # 565352) revealed that the product was processed according to validated specification requirements and parameters.Smith medical periodically monitors supplied drug components stability post ethylene oxide sterilization throughout their expiry.Our records show that the supplied drug components continue to meet potency specifications.Complaint information review found no trends relevant to the drug lot number, reported in this complaint.The anesthetics are supplied items and the complaint notification was forwarded to the supplier.Based on available information and evidences no product quality problem could be confirmed.Possible causes of lack of effect may be administration technique or patient anatomical variations, including pathological or psychological factors.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX® SINGLE SHOT EPIDURAL TRAYS
Type of Device
ANESTHESIA CONDUCTION KIT
Manufacturer (Section D)
SMITHS MEDICAL ASD; INC.
6000 nathan lane n
minneapolis, MN 55442
MDR Report Key8323738
MDR Text Key135613185
Report Number3012307300-2019-00476
Device Sequence Number1
Product Code CAZ
UDI-Device Identifier15019517115697
UDI-Public15019517115697
Combination Product (y/n)N
PMA/PMN Number
E332605
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 04/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/08/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/01/2020
Device Catalogue NumberNEPI-NLD-15597-20
Device Lot Number3669767
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/25/2019
Date Manufacturer Received03/15/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Congenital Anomaly; Required Intervention; Disability;
-
-