Catalog Number 8065990713
Device Problem Improper or Incorrect Procedure or Method (2017)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/16/2019
Event Type  Injury  
Manufacturer Narrative
The device history records (dhr) for the device were reviewed.The associated device was released based on acceptance criteria.The root cause is user error: the surgeon reported that he entered the refraction data of the right eye on the left eye by mistake and proceeded with the laser shooting.The system would give a notification message but the surgeon can still overcome to shoot the laser as the system proceed with the data entered by the user.A product malfunction is excluded.(b)(4).
 
Event Description
A surgeon reported that the manifest refraction data of the right eye was entered for the left eye's laser treatment.There is a high difference in cylinder between the eyes.The surgeon will wait for one to three months postoperatively to decide how to proceed.Additional information requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WAVELIGHT EX500 EXCIMER LASER
Type of Device
OPHTHALMIC EXCIMER LASER SYSTEM
Manufacturer (Section D)
WAVELIGHT GMBH
am wolfsmantel 5
erlangen 91058
GM  91058
Manufacturer (Section G)
WAVELIGHT GMBH
am wolfsmantel 5
erlangen 91058
GM   91058
Manufacturer Contact
cindy milam
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152231
MDR Report Key8331060
MDR Text Key135873648
Report Number3003288808-2019-00153
Device Sequence Number1
Product Code LZS
Combination Product (y/n)N
Reporter Country CodeJO
PMA/PMN Number
P020050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/12/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number8065990713
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/16/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/26/2017
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-