• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON SPINAL SYSTEM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT SOFAMOR DANEK PUERTO RICO MFG CD HORIZON SPINAL SYSTEM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Catalog Number 55410007550
Device Problem Scratched Material (3020)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/29/2019
Event Type  malfunction  
Manufacturer Narrative
Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient underwent deformation correction surgery at th4-l5.Intra-op, the set screw overturned, which destroyed the fixed-angle screw.The damaged screw was then explanted and replaced with a multi-axial screw.The set screw overturned because the surgeon did not use a counter-torque instrument.There were no patient symptoms or complications as a result of this event.
 
Manufacturer Narrative
Product analysis: observation: after visual and optical examination, it appears that threads of the set screw are visibly damaged.The screw head appears to function properly when tested with a sample screw.Conclusion: this is consistent with misalignment of the set screw threads during construct assembly.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD HORIZON SPINAL SYSTEM
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
Manufacturer (Section G)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
Manufacturer Contact
stacie ziemba
1800 pyramid place
memphis, TN 38132
9013963133
MDR Report Key8346832
MDR Text Key136853212
Report Number1030489-2019-00160
Device Sequence Number1
Product Code NKB
UDI-Device Identifier00643169074187
UDI-Public00643169074187
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K113174
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 05/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/18/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number55410007550
Device Lot NumberH5247983
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/29/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/29/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/20/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age13 YR
-
-