Catalog Number 55410007550 |
Device Problem
Scratched Material (3020)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/29/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the patient underwent deformation correction surgery at th4-l5.Intra-op, the set screw overturned, which destroyed the fixed-angle screw.The damaged screw was then explanted and replaced with a multi-axial screw.The set screw overturned because the surgeon did not use a counter-torque instrument.There were no patient symptoms or complications as a result of this event.
|
|
Manufacturer Narrative
|
Product analysis: observation: after visual and optical examination, it appears that threads of the set screw are visibly damaged.The screw head appears to function properly when tested with a sample screw.Conclusion: this is consistent with misalignment of the set screw threads during construct assembly.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|