Failure analysis was performed on the main board.Visual inspection: no anomalies.Benchtop analysis: assembled into a golden unit.Ran on automated test console.Failed atrial pulse noise test, 169.12mv.Measured atrial pulse noise is within the requirements of the atrial pulse noise test (0-100 mv).Measured atrial pulse noise on the bench at 50.0mv.Logs analyzed, no anomalies.Confirmed unit fails atrial pulse noise test, but atrial pulse noise level measured is within device specifications.Conclusion: no defect found.If information is provided in the future, a supplemental report will be issued.
|