The c-port xa (fg-000100) and c-port flexa (fg-000150) share the same design in respect to activation of clamping, deployment, and the energy source used to accomplish these actions.Thus, the same malfunction is noted in the two devices.Corrected data: device problem codes were added.
|
The c-port device was removed from the packaging by the technician and passed to the surgeon to prepare.The graft was loaded onto the device and the device was activated.When the deployment button was pushed, the device did not deploy.The device was then removed from the surgical field and placed on a non-sterile back table.Approximately 10 to 15 minutes later,the device broke into pieces while on the non-sterile back table, accompanied by a loud noise.There was no reported adverse event either to the patient or to surgical personnel.
|