Catalog Number 5MAXACE068KIT |
Device Problems
Break (1069); Device Damaged Prior to Use (2284)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 01/10/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
|
|
Event Description
|
During preparation for a medical procedure using a penumbra system ace 68 hi-flow kit, the penumbra system ace 68 reperfusion catheter (ace68) was found to be broken at the proximal end upon removal from the kit packaging.The damage to the ace68 was found prior to use and therefore it was not used in the procedure.The procedure was completed using another ace68 from a new kit.
|
|
Event Description
|
During preparation for a medical procedure using a penumbra system ace 68 reperfusion catheter (ace68) from a penumbra system ace 68 hi-flow kit, and a penumbra system 3max reperfusion catheter, the ace68 was found to be broken at the proximal end upon removal from the kit packaging.The damage to the ace68 was found prior to use and, therefore, it was not used in the procedure.The physician then began using a 3maxc and another ace68 from a new kit to continue the procedure.A report was received on 07-mar-2019 which indicated that after advancement of the ace68 over the 3maxc into the m1 segment of the middle cerebral artery (mca), the physician retracted the 3maxc to remove it and reported that the proximal part of the catheter broke as it was passing through the rotating hemostasis valve (rhv).The 3maxc was then completely removed and the procedure was completed successfully using the same ace68.There was no report of an adverse effect to the patient.
|
|
Manufacturer Narrative
|
Results: the ace68 was fractured approximately 57.5 cm from the hub.Conclusions: evaluation of the returned ace68 confirmed a fracture.If the device is retracted out of its packaging tray without removing the top portion of the tray, the ace68 may become wedged between the two trays and result in a fracture.Evaluation of the returned 3maxc confirmed a fracture.If the device was forcefully retracted through a tightened rhv or at an extreme angle through the rhv, damage such as a fracture may occur.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
|
Search Alerts/Recalls
|