If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.This report is for an unknown device/unknown lot.Part and lot number are unknown; udi number is unknown.
|
It was reported that tibial bone trunk was fractured during tka surgery by using the attune revision pr tibial tray (p/n: unknown) on (b)(6) 2019.The cemented reamer tip interfered with the proximal tibial bone trunk and the tibia was fractured.The fracture was found by ct after surgery.The patient used steroids, so bone quality was bad, and it was quite fragile.It was considered that it may have interfered with the distal cortical bone because the cemented reamer was a little big.No further information is available.
|
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).No device associated with this report was received for examination.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : null.Device history batch : null.Device history review : null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|