Catalog Number 8065990713 |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/15/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device history records (dhr) for the device was reviewed.The associated device was released based on company acceptance criteria.Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
|
|
Event Description
|
During maintenance, an arf leakage was noted inside the laser head.No impact or harm.
|
|
Manufacturer Narrative
|
The field service engineer (fse) placed the order for laser head and replaced the laser head, focus lens, beam splitter, main deflector and performed service installation record.No deeper root cause analysis is possible.The possible root cause is a worn part of the laser head.The manufacturer internal reference number is: (b)(4).
|
|
Manufacturer Narrative
|
According to serial number of laser head, this laser head has a c-ring sealing inside the laser head.The root cause is the leakage of the c-ring sealing inside laserhead.The supplier has improved the sealings and implemented o-ring sealings.The root cause is the leakage of the c-ring inside laserhead.The manufacturer internal reference number is: (b)(4).
|
|