Catalog Number 8065990713
Device Problem Leak/Splash (1354)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/15/2018
Event Type  malfunction  
Manufacturer Narrative
The device history records (dhr) for the device was reviewed.The associated device was released based on company acceptance criteria.Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
During maintenance, an arf leakage was noted inside the laser head.No impact or harm.
 
Manufacturer Narrative
The field service engineer (fse) placed the order for laser head and replaced the laser head, focus lens, beam splitter, main deflector and performed service installation record.No deeper root cause analysis is possible.The possible root cause is a worn part of the laser head.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
According to serial number of laser head, this laser head has a c-ring sealing inside the laser head.The root cause is the leakage of the c-ring sealing inside laserhead.The supplier has improved the sealings and implemented o-ring sealings.The root cause is the leakage of the c-ring inside laserhead.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WAVELIGHT EX500 EXCIMER LASER
Type of Device
OPHTHALMIC EXCIMER LASER SYSTEM
Manufacturer (Section D)
WAVELIGHT GMBH
am wolfsmantel 5
erlangen 91058
GM  91058
MDR Report Key8358581
MDR Text Key137143341
Report Number3003288808-2019-00205
Device Sequence Number1
Product Code LZS
Combination Product (y/n)N
PMA/PMN Number
P020050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup,Followup
Report Date 06/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/21/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065990713
Was Device Available for Evaluation? Yes
Date Manufacturer Received05/26/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-