• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® I.C. FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® I.C. FOLEY CATHETER Back to Search Results
Model Number 0165SI18
Device Problems Unsealed Device Packaging (1444); Tear, Rip or Hole in Device Packaging (2385)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.
 
Event Description
It was reported that the foley catheter was in an unsealed package.
 
Event Description
It was reported that the foley catheter was in an unsealed package.
 
Manufacturer Narrative
The reported event was confirmed, but the cause was unknown.Visual evaluation of the returned sample noted one bardex latex foley in opened original packaging.The packaging on the end of the opening appeared to be cut.The blue sleeve and the strip pack were both torn, exposing the catheter.Based on the evaluation, it could not be determined at what point the packaging was damaged before it was received by patient.Holes/ broken/ open seals are out of specification.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "this is a single use device.Do not resterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and /or lead to injury, illness or death of the patient.Valve type: use luer slip tip syringe.Do not use needle.Do not exceed recommended capacities as stated on the inflation lumen of the catheter.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Caution: do not aspirate urine through drainage funnel.Warning: on catheter, do not use ointments or lubricants having a petrolatum base.They will damage latex and may cause balloon to burst.After use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable local, state and federal laws and regulations.Caution: this product contains natural rubber latex which may cause allergic reactions.Sterile unless package is opened or damaged.Do not use if package is damaged.Recommended indwelling time not to exceed 28 days.Bard, bardex, and the i.C.Logo are trademarks and/or registered trademarks of c.R.Bard, inc.*bacti-guard® silver alloy coating is licensed from bactiguard ab.Bacti-guard is a registered trademark of bactiguard ab.©2014 c.R.Bard, inc.All rights reserved.Catheter made in u.S.A." section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® I.C. FOLEY CATHETER
Type of Device
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key8359224
MDR Text Key137125810
Report Number1018233-2019-00933
Device Sequence Number1
Product Code MJC
UDI-Device Identifier00801741016769
UDI-Public(01)00801741016769
Combination Product (y/n)N
PMA/PMN Number
K910318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup
Report Date 06/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/21/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/31/2019
Device Model Number0165SI18
Device Catalogue Number0165SI18
Device Lot NumberNGAP0795
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/31/2019
Date Manufacturer Received05/28/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-