MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SYSTEM ASPIRATION PUMP MAX 110; DXE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SYSTEM ASPIRATION PUMP MAX 110; DXE Back to Search Results
Catalog Number PMX110
Device Problems Misconnection (1399); Use of Device Problem (1670); Device Contamination with Body Fluid (2317)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/23/2019
Event Type  malfunction  
Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
 
Event Description
The patient was undergoing a thrombectomy procedure in the proximal left anterior descending artery (lad) using a penumbra system aspiration pump max 110 (pump max).During the procedure, the physician noticed that blood was entering the pump max as it was aspirating.The physician then found that the indigo system catrx aspiration catheter (catrx) was connected incorrectly to the pump max, allowing for blood to enter.The catrx and connection tubes were then re-adjusted and connected correctly to the pump max, and the procedure resumed.The procedure was completed using the same catrx and the same pump max.There was no report of an adverse effect to the patient.
 
Manufacturer Narrative
Results: dried blood was found on the pump housing and inside the vacuum inlet.Conclusions: evaluation of the returned pump max confirmed blood was aspirated inside the vacuum pump.The reported complaint indicated that the aspiration tubing was directly connected to the pump vacuum inlet instead of the canister supplied by penumbra and blood was subsequently aspirated into the pump.If the fluid enters the vacuum pump assembly, the pump may not function properly.Penumbra pumps are visually inspected and functionally tested during incoming quality inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM ASPIRATION PUMP MAX 110
Type of Device
DXE
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key8363387
MDR Text Key136994510
Report Number3005168196-2019-00337
Device Sequence Number1
Product Code DXE
UDI-Device Identifier00814548012773
UDI-Public00814548012773
Combination Product (y/n)Y
PMA/PMN Number
K160533
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 01/01/2005,01/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/22/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberPMX110
Device Lot NumberF25844-32
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2019
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Date Manufacturer Received04/03/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age30 YR
-
-