Model Number G03793
Device Problem Retraction Problem (1536)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/31/2019
Event Type  malfunction  
Manufacturer Narrative
Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.Prior to distribution, all echotip ultra fiducial needles are subjected to a visual inspection to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event Description
During an endoscopic ultrasonography (eus), the physician used a cook echotip ultra fiducial needle.After the deployment of the third fiducial marker, the needle was unable to retract to point zero (0) in length.The fourth fiducial marker could not be deployed and remained stuck in the needle.This caused the deployment of three (3) fiducial markers only.The needle was cautiously removed from the endoscope.Other than the deployed fiducials, a section of the device did not remain in the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHOTIP ULTRA FIDUCIAL NEEDLE
Type of Device
NEU, MARKER, RADIOGRAPHIC, IMPLANTABLE, BIOPSY NEEDLE KIT
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
scottie fariole
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key8367141
MDR Text Key141066577
Report Number1037905-2019-00090
Device Sequence Number1
Product Code NEU
UDI-Device Identifier00827002037939
UDI-Public(01)00827002037939(17)190822(10)W4107329
Combination Product (y/n)N
Reporter Country CodeHK
PMA/PMN Number
K141356
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial
Report Date 01/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/25/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/22/2019
Device Model NumberG03793
Device Catalogue NumberECHO-22-F
Device Lot NumberW4107329
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/31/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/22/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age59 YR
Patient Weight68
-
-