• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION SWIFT-LOCK ANCHOR; SCS ANCHOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION SWIFT-LOCK ANCHOR; SCS ANCHOR Back to Search Results
Model Number 1192
Device Problem Material Split, Cut or Torn (4008)
Patient Problem Inadequate Pain Relief (2388)
Event Date 02/06/2019
Event Type  Injury  
Event Description
Device 4 of 4.Reference mfr.Report #: 3006705815-2019-00573.Reference mfr.Report #: 1627487-2019-02275.Reference mfr.Report #: 3006705815-2019-00574.It was reported the patient¿s leads migrated, as verified via x-rays.Diagnostics revealed low impedances on both of the patient¿s leads.To address the issue, the physician opted to perform surgical intervention on (b)(6) 2019 wherein the midline incision was reopened, and it was then discovered both of the anchors had torn.The physician therefore replaced both leads and both anchors with new models.Impedances were then checked and showed within normal limits on every contact.Postoperatively, effective therapy was restored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWIFT-LOCK ANCHOR
Type of Device
SCS ANCHOR
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
andrea deitz
6901 preston road
plano, TX 75024
9723098000
MDR Report Key8368659
MDR Text Key137111053
Report Number1627487-2019-02277
Device Sequence Number1
Product Code GZB
UDI-Device Identifier05415067024084
UDI-Public05415067024084
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K092371
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 02/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/25/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/14/2020
Device Model Number1192
Device Lot Number6482288
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/08/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/06/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/15/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Weight68
-
-