Catalog Number 950502071
Device Problems Difficult to Insert (1316); Positioning Problem (3009)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/11/2019
Event Type  malfunction  
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device is an instrument and is not implanted/explanted.
 
Event Description
It was reported that the balancing bolt is stripped threads and pindriver does not seat in power chuck.No surgical delay.
 
Manufacturer Narrative
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary : examination and functional evaluation of the returned device could not replicate the reported event.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : null.Device history batch : null.Device history review : null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HP POWER PIN DRIVER
Type of Device
KNEE INSTRUMENT : INSERTION DEVICES
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8371358
MDR Text Key137297886
Report Number1818910-2019-85824
Device Sequence Number1
Product Code HWE
UDI-Device Identifier10603295227878
UDI-Public10603295227878
Combination Product (y/n)N
PMA/PMN Number
UNK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup,Followup
Report Date 02/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/26/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number950502071
Device Lot NumberNB82839
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/04/2019
Date Manufacturer Received04/05/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-