RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Overheating of Device (1437)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2019.Date of report: 26feb2019.The customer could not duplicate the reported issue.The customer stated it could have been a settings issue or more likely patient error.The unit was ran in biomed for 3 to 4 hours with air intake covered.Filters were clean, units were put back into service and a week later still no errors.The unit successfully passed the required performance verification test.
|
|
Event Description
|
The customer reported error blower temperature high (e/c 1122).The unit was in clinical use at the time the reported issue was discovered; however, there was no harm to the patient or the user.Event date not specified; estimate used.
|
|